Data-Logging; Alarms - Consort EV202 Anleitung

Elektrophoresenetzgerät
Vorschau ausblenden Andere Handbücher für EV202:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1. Select SET-UP and press SET.
2. Select DATA and press SET.
3. Select DURATION and press SET.
4. Select the desired time for the data-logging and press SET.
5. Select LOG and press SET.
6. Decide whether or not all data (voltage, current, power, time) during
the next experiments should be logged and press SET.
7. Select SET-UP and press SET to return.
Starting a new experiment will erase the previous data in memory.
Stored data can be transferred to a computer via the RS232 output using
free available software (www.consort.be).
At any time, you can return in the stand-by mode by holding SET pressed
during 2 seconds.
GROUND LEAKAGE
A dangerous ground leakage is present (check thoroughly the complete
assembly). Your power supply is NOT defective! See Frequently Asked
Questions at the end of this manual.
SHORT-CIRCUIT
Connected tank is short-circuited or has a very low resistance (check
thoroughly the complete assembly). Your power supply is NOT defective!
See Frequently Asked Questions at the end of this manual.
OVERLOAD
Connected tank has a too low resistance (check specifications). Your
power supply is NOT defective! See Frequently Asked Questions at the
end of this manual.
CURRENT<1mA
The output current is lower than 1 mA (switch the alarm off in the SET-
UP menu). Your power supply is NOT defective! See Frequently Asked
Questions at the end of this manual.
CALL SERVICE
Failure in the electronics. See Support on www.consort.be.
You could also try a general RESET: switch the instrument on while holding
SET pressed. Attention! All stored programs and data will be erased!
6

Data-logging

---- STAND-BY ----
SET-UP
---- SET-UP ----
DATA
----- DATA -----
DURATION
LOG DATA
FOR:
----- DATA -----
LOG
LOG
DATA:
----- DATA -----
SET-UP

Alarms

*** ERROR ***
GROUND LEAKAGE
*** ERROR ***
SHORT-CIRCUIT
*** ERROR ***
OVERLOAD
*** ERROR ***
CURRENT<1mA
*** ERROR ***
CALL SERVICE
10h
YES