1. Reset the power supply by switching on while holding SET pressed.
2. Switch the power supply off.
3. Switch the power supply on while holding RUN pressed.
4. Adjust the contrast of the display and press SET or RUN.
5. Short circuit the output and press RUN. The uncalibrated output
values are shown for a while.
6. Remove the short circuit, leave the output open and press RUN. The
uncalibrated output values are shown for a while.
7. The zero points of all output values are now calibrated.
For software versions <1.2 (EV245: 1.4) the sequence of points 5. and 6. are
inverted.
WARRANTY CERTIFICATE
This instrument (excluding all accessories) is warranted against defective material
and workmanship for a period of thirty-six (36) months from the date of shipment ex
factory.
Accessories and breakable items such as electrodes are not warranted unless proven
to be defective before shipment.
The original purchase order numbers, Consort invoice numbers and serial numbers of
the products must be provided.
CONSORT will repair all defective equipment returned to it during the warranty
period without charge (CIF Turnhout prepaid by sender), provided the equipment has
been used under normal laboratory conditions and in accordance with the operating
limitations and maintenance procedures described in the instruction manual and when
not having been subject to accident, alteration, misuse or abuse.
If the products have been used with or have come into contact with fluids, an MSDS
(material safety data sheet) must be supplied prior to issuing a return authorisation.
A return authorisation must be obtained from CONSORT before returning any product
for warranty repair on a freight prepaid basis!
26

EV2xx Calibration

SYSTEM
RESET?
SET=Yes
STOP=No
ENTER
CONTRAST:
Short circuit
output!
RUN=OK
Please wait!
4V
3mA
Leave output
open!
RUN=OK
Please wait!
4V
3mA
END CALIBRATION
2
0W
0W

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Consort EV202

Inhaltsverzeichnis