Herunterladen Inhalt Inhalt Diese Seite drucken

Emc Information - ULAIZER HOME CN-02 MY Benutzerhandbuch

Kompressor vernebler
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 60
EN
28
Guidance and manufacturer's declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that is used in such an environment.
Emissions
RF emissions CISPR 11
RP emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Recommended separation distances between portable and mobile RF communications equip-
ment and the IH25/1
The IH25/1 is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the IH25/1 can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the IH25/1 as recommended below, according to the maximum output
power of the communications equipment.
Separation distance according to frequency of transmitter m
Rated maximum
output power of
150 kHz to 80 MHz
transmitter W
d = 1.2√P
0.01
0.12
0.1
0.38
1
1.2
10
3,8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Notes:
1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.

10. EMC INFORMATION

Compliance
Electromagnetic environment—guidance
Group 1
The device uses RF energy only for its in-
ternal function. Therefore, its RF emissions
are very low and not likely to cause any in-
terference in nearby electronic equipment.
Class B
The device is suitable for use in all estab-
lishments, including domestic establish-
Class A
ments and those directly connected to the
public low-voltage power supply network
Complies
that supplies buildings used for domestic
purposes.
80 to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
800 MHz to 2,5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis