Using Setup in pH Mode
Clear
4 7
setup
mode
standardize
Set
2 4 7 10 12
setup
mode
standardize
The setup key lets you clear all buffers/standards
that have been entered, review standardizing
information, and select the desired buffer or set
of standards.
Note: Press setup repeatedly to return to
measurement mode.
1. Press setup; the word Clear flashes on the
display. Continue with this step only if you wish
pH
to clear all buffers entered. To clear all existing
buffers, press enter. The meter clears all buffers
and returns to measuring mode.
2. Press setup again to show electrode perform-
ance. If two buffers are stored, OK is displayed
with the electrode symbol, the slope between
the first and second buffers, and the two buffer
symbols.
enter
3. Press setup again to show the electrode slope
between the second and third buffers (if three
buffers have been entered) and the second and
third buffer symbols.
Note: The symbols are displayed as described
only if buffers are stored.
4. Press setup again; Set flashes and the first
buffer set (U.S.: buffer series) is shown.
pH
5. Press enter to select the set of buffers displayed
or
Press setup again to show the next set of
buffers (Merck buffer set), and again for the
third buffer set (NIST buffer set).
enter
13