Herunterladen Inhalt Inhalt Diese Seite drucken

SciCan Bravo Serie Bedienerhandbuch Seite 72

Inhaltsverzeichnis

Werbung

Ausdruck nach Alarm
Model
BRAVO 17V
S/N
12 JP 0001
Ver. SW
Exxxx/JPyyyyyy
Counter
0007~001
Selection
134 °C POROUS/WRAPPED
Temperature
134 °C
Pressure
2.10 Bar
Process time
4 min
Stand-by
HIGH
Pre-vacuum
FRACTIONATED
Drying
STANDARD
CYCLE START
01/02/11
11:30
Time
T1
P
T2
-------------------------------------------------------------------
00:01
CS
075.1 -0.00
130.9
00:11
074.9 -0.28
133.3
00:21
074.4 -0.46
146.3
00:31
074.3 -0.57
152.6
00:35 ..
074.3 -0.59
154.2
00:51
..
078.9 -0.62
152.2
01:01 ..
074.9 -0.73
146.6
01:27
..
047.8 -0.78
149.3
01:57
..
047.8 -0.80
155.3
02:07 ..
076.5 -0.57
149.9
02:17
..
081.1 -0.49
142.1
...............
...............
08:15
...
068.4 -0.76
151.8
08:22 ...
061.1 -0.80
153.6
08:32 ...
097.4 +0.01
154.7
08:42
...
104.6 +0.24
148.9
...............
...............
15:04
...
135.5 +2.15
143.3
15:19
...
135.9 +2.17
148.5
15:28
...
135.3 +2.16
153.6
...............
...............
19:19
...
135.5 +2.15
157.4
19:34
...
134.4 +1.07
157.0
19:49
...
108.3 +0.25
156.4
19:53
DS
104.4 +0.00
156.1
STERILISATION
NEGATIVE
ALARM CODE:
A112
DESCRIPTION
PTC SHORTCIRCUIT
CAUTION !
PLEASE REFER TO USER MANUAL
68
68
Ausdruck des Programms
HELIX/BD-TEST
Model
S/N
Ver. SW
Counter
Selection
Temperature
Pressure
Process time
CYCLE START
Time
----------------------------------------------------------
00:01
CS
02:06
1PV
04:35
1PP
T3
T4
05:45
2PV
07:02
2PP
115.2
093.4
08:15
3PV
114.2
094.0
11:00
..
113.2
094.5
11:14
..
112.2
095.0
12:14
111.9
095.2
13:14
110.4
095.6
14:14
109.6
095.7
14:45
..
107.7
095.7
15:20
..
105.8
095.4
16:34
...
105.2
095.1
18:21
...
104.6
094.6
19:21
..
20:06
CE
104.7
102.3
12:33
MAX
104.5
101.7
14:44
MIN
104.0
100.8
Drying pulses
103.7
101.0
CYCLE END
111.7
131.7
HELIX TEST COMPLETE
Please attach the indicator hereunder
113.5
132.6
115.9
133.0
126.5
132.5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
126.8
131.2
126.8
119.9
126.6
116.2
SOLLTE DER USB-STIFT NICHT EINGESTECKT SEIN, BESTEHT IMMER DIE
MÖGLICHKEIT DER ELEKTRONISCHEN ARCHIVIERUNG DER DRUCKBE-
RICHTE
BRAVO 17V
12 JP 0001
Exxxx/JPyyyyyy
0011/0019
HELIX TEST
134 °C
2.10 bar
3.5 min
01/02/11
16:38
C
bar
076.4
+0.00
089.3
-0.89
120.4
+0.99
062.5
-0.78
120.2
+0.97
061.1
-0.79
135.6
+2.15
136.0
+2.17
135.6
+2.14
135.6
+2.15
135.5
+2.14
135.4
+2.14
111.5
+0.00
047.8
-0.89
059.5
-0.86
075.4
-0.50
078.7
-0.04
136.0
135.4
01
01/02/11
17:01
OPERATOR
ANMERKUNG
Ausdruck des
ProgrammsVAKUUMTEST
Model
BRAVO 17V
S/N
12 JP 0001
Ver. SW
Exxxx/JPyyyyyy
Counter
0011/0019
Selection
VACUUM TEST
CYCLE START
01/02/11
11:37
Time
C
bar
----------------------------------------------------------
00:00
CS
035.0
+0.00
01:39
E1F
037.4
-0.80
6:39
E2F
038.4
-0.79
22:39
E3F
042.0
-0.79
23:54
CE
045.5
-0.01
CYCLE END
01/02/11
12:01
VACUUM TEST:
POSITIVE
OPERATOR
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für SciCan Bravo Serie

Diese Anleitung auch für:

Bravo 17vBravo 17Bravo 21v

Inhaltsverzeichnis