3. Operation
ment off by pointing the unit to the target to recall the last value when you switch
the unit on.
Select Average (AVG), Minimum (MIN) and Maximum (MAX) values:
Press the (Min/Max] key until the appropriate message is displayed. The unit will
refresh the values of each measurement from the power on. To reset the values,
press the [2nd] key or switch the unit off and on.
Emissivity adjustments
Press in sequence the [2nd] and [Set e] keys to change the emissivity from 0.30
to 1.00 with 0.01resolution. On the lower display the set value is shown. Press
the ['] and [-] keys, to increase and decrease the emissivity. Press the [Enter]
key to store and return on the normal mode measurement.
Set low (LAL) and high (HAL) alarms
Press in sequence the [2nd] and [HI] keys to change the High alarm set point.
„HAL" will be seen on the display. The stored High alarm will be displayed. The
High alarm level is to be set by adjusting the ['] or [-] keys. Press the [HI] key
to enable/disable the acoustic alarm. Press the [Enter] key to store the displayed
value and return in normal mode.
Press the sequence the [2nd] and [LO] keys to change the Low alarm set point.
„LAL" will be seen on the display. The LOW alarm level is to be set by adjusting
the ['] or [-] keys. Press the [LO] key to enable/disable the acoustic alarm.
Press the [Enter] key to store the displayed value and return in normal operation
14
Hotline: +49 29 53 / 73 211
Fax: +49 29 53 / 73 250
IR-Temp 100L