Herunterladen Inhalt Inhalt Diese Seite drucken

Lovibond Tintometer MD100 Bedienungsanleitung Seite 43

Photometer-system
Vorschau ausblenden Andere Handbücher für Tintometer MD100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Methods
GB
Notes:
1. Vial cleaning:
As many household cleaners (e.g. dishwasher detergent) contain reducing substances,
the subsequent determination of Bromine may show lower results. To avoid any meas-
urement errors, only use glassware free of Chlorine demand.
Preparation: Put all applicable glassware into Sodium hypochlorite solution (0.1 g/l) for
one hour, then rinse all glassware thoroughly with deionised water.
2. Preparing the sample:
When preparing the sample, the lost of Bromine, e.g. by pipetting or shaking, must be
avoided. The analysis must take place immediately after taking the sample.
3. The DPD colour development is carried out at a pH value of 6.2 to 6.5. The reagents
therefore contain a buffer for the pH adjustment.
Strong alkaline or acidic water samples must be adjusted between pH 6 and pH 7 before
the reagent is added (use 0.5 mol/l Sulfuric acid resp. 1 mol/l Sodium hydroxide).
4. Exceeding the measuring range:
Concentrations above 22 mg/l Bromine can lead to results showing 0 mg/l. In this case,
the water sample must be diluted with water free of Chlorine and the measurement
repeated.
5. Oxidising agents such as Bromine, Ozone etc. interfere as they react in the same way as
Bromine.
Reagent
DPD No. 1
MD100_10b 05/2019
Form of reagent/Quantity
Tablet / 100
Order-No.
511050BT
43

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis