Control Test
Range 80-120
mg/dL
Out of range results may be caused by:
when you are not comply the instructions on this USER'S
GUIDE,
expired, contaminated or watered-down control solution,
expired or damaged test strip,
control solution test is done out of temperature range 20°C
to 25°C,
or a problem on the
If your control solution test results are out of the range
continuously, do not use
the control solution. Please contact the local distributor.
STEP 8
Roll the eject wheel to discard the used strip. Test result is
stored in memory automatically.
CONTROL SOLUTION TESTING
STEP 6
Beeps when test is fi nished. The test
result of control solution is displayed on
screen.
STEP 7
Compare the test result with control
solution range which is printed on test
strip container. If the test result is not
within the range, the
strip may worked improperly. Repeat the
control solution test.
2in1.® BASIC
2in1.®
BASIC, the test strips and/or
2in1.® BASIC
and
25