4. Specifications

The ND 130x QUADRA-CHEK is an advanced evaluation unit for performing 2, 3 and 4
axis measurements at very high levels of precision and accuracy. This unit is designed for
indoor use only. The ND 130x components shall only be installed as described in these
instructions. Mounting, installation and maintenance are to be performed by qualified
personnel only.
Power input:
AC 100 V ... 240 V (–15 % ... +10 %)
50 Hz ... 60 Hz (±3 Hz)
(max. 100 W)
Replaceable fuse:
T 1.6 A / 250 V, 5 mm x 20 mm
Quantity: 2
Power consumption
X, Y 13 :
max. 750 mW
Z, Q 13 :
max. 750 mW
CNC 9 :
max. 13 W
Light 6 :
rated 800 mW (100 mW per lamp)
Environmental conditions
Operating temperature:
0 °C ... 45 °C
Storage temperature:
– 20 °C ... 70 °C
Relative humidity:
† 80 %
Altitude:
† 2000 m
Protection:
IP00, front IP40
Overvoltage category:
II, equipment intended to be supplied from the building wiring
Pollution degree:
2
Mass:
ND with stand: approx. 4.8 kg
ND with mounting adapter: approx. 2 kg

5. Mounting

The following items are included with the product:
• Power cord
• Stand or mounting adapter
• 2-D demo part
• ND 1300 QUADRA-CHEK Installation Instructions
The following accessories are available for this product:
• Calibration standard, ID 681047-01
• 2-D demo part, ID 681047-02
• Foot switch, ID 681041-01
• Fiber-optic cable, ID 681049-xx
• Holder, ID 681050-xx
• Fiber-optic cable connector, ID 681049-xx
• Stand, ID 382892-02
• Mounting adapter, ID 682419-01
• Joystick, ID 681044-xx
Refer to " Accessories" on page 116 for additional information. Refer to the instructions
provided with an accessory for additional information.
Do not mount in a position that makes it difficult to access the power switch or
power cable.