Herunterladen Diese Seite drucken

ACCU-CHECK Rapid-D Link Gebrauchsanweisung Seite 11

Infusion set

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 15
Fill volumes of the complete set:
30 cm/12"
60 cm/24"
80 cm/31"
110 cm/43"
Picture 4
Hold the Accu-Chek Rapid-D Link firmly between thumb and forefinger at the disc holder
(E). With the other hand, remove the adhesive backing (D) from the self-adhesive (C) and
twist the protective cover (B) off the steel cannula (A).
Picture 5
With your free hand, pinch together a bit of subcutaneous tissue at the disinfected site.
Insert the Accu-Chek Rapid-D Link under the skin at a perpendicular angle. Press the self-
adhesive (C) firmly onto the skin in order to ensure secure seating of Accu-Chek Rapid-D
Link. Bend the disc holder (E) slightly sideways and remove it. In doing this, make sure that
the self-adhesive comes to lay flat on the skin.
WARNING
Change the head-set every 2 days (max. 48 hours). Otherwise you risk insulin resistance
or inflammation of the infusion site.
Never use the transfer-set for more than 6 days.
Picture 6
Form a loop to relieve stress and secure it into place with a strip of adhesive plaster. Start
distribution through the pump.
approx. 7 units U 100 insulin
approx. 10 units U 100 insulin
approx. 13 units U 100 insulin
approx. 18 units U 100 insulin
approx. 70 μl
en
approx. 100 μl
approx. 130 μl
approx. 180 μl
9