Herunterladen Diese Seite drucken

Sartorius Stedim SARTOFLOW Alpha plus SU Bedienungsanleitung Seite 5

Installation der sartocon slice self contained und montage des vorsterilisierten crossflow bag loops

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Installation of the
Sartocon Slice Self Contained
Before the assembly and cleaning of the filter holder you must carefully read
the operating manual of the Sartocon Slice Self Contained Unit! In the case of
incorrect handling of the filter cassettes severe injuries may result or the filter
cassette may be damaged.
You must only use the provided torque wrench or the hydraulic assembly
to clamp the Self Contained Unit!
- Incorrect handling may result cassette damage.
Keep the stud bolts and screw nuts clean and free of grease!
Greased or polluted threads lead to wrong clamp forces which may destroy
the cassettes.
For the installation or change of the filter cassettes follow the instructions
in „Sartocon Slice Self Contained Unit "
of the chapter.
1. Install the Self Contained unit in the filter holder and clamp it with the
recommended torque, or with the appropriate hydraulic pressure.
Figure 7
– Tighten the Cassette with the recommended clamping force according to the
selected clamping device.
– When installing the Cassette make sure that the grooves are in correct orientation
(to accommodate the upper and lower rod sites).
– Keep the threads of the rods clean (avoid grease).
Do not forget hydraulic clamping the upper locking bars | rods
Clamping force for Sartocon
Sartocon
®
Slice Self Contained Holder
®
Sartocon
Slice Self Contained
[1]
, see bibliography at the end
®
Slice Self Contained
Clamping force
Operating parameters
14–17 kN
Installation and Starting Up
Clamping force
Operating parameters
Torque wrench
20 Nm
7