Problem
LCD shows low battery
symbol
LCD shows "Er 0"
LCD shows "Er 1"
LCD shows "Er 2"
LCD shows "Er 3"
LCD shows "Er 4"
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er A"
No response when you press
button or install batterypiles
Explanation
Low Battery
Pressure system is unstable before
measuremente
Fail to detect systolic pressure
Fail to detect diastolic pressure
Pneumatic system blocked or cu is too
tight during in ation
Pneumatic system leakage or cu is too
loose during in ation
Cu pressure above 300mmHg
More than 3 minutes with cu pressure
above 15 mmHg
EEPROM accessing error
Device parameter checking error
Pressure sensor parameter erroron
Incorrect operation or strong
electromagnetic interference.
Solution
Change all the batteries
Try again without moving
Apply the cu correctly
and try again
Measure again after 5
minutes. If the error is still
there, please contact your
local iHealth distributor
Take out the batteries, wait
ve minutes and then
reinstall batteries