6.0 Interpreting results

The unit has an operating range of 0.50 mL/min to 500 mL/min (Figure 7). If the flow is less than
0.50 mL/min., the display will read "under range."
If the flow exceeds 515 mL/min, the display will read "over range." Excessively high flow rates may dam-
age this unit.
NOTE: units of mL/min are equivalent to ccm.

6.1 Flow range display

The unit automatically adjusts the resolution of the display depending on the flow range being measured.
Table III shows the resolution of the flow ranges.
Table III:
Display resolution vs. flow range.
Flow range
0.50 – 9.99
10.0 – 99.9
100 – 500
www.restek.com
Display resolution (mL/min)
0.01
0.1
1
- 6 -
Figure 7:
Description of flow ranges.
Flow Range
Display reads
Display reads
"under range"
flow values
0.50 mL/min
Figure 6:
Example flow
value.
Display reads
"over range"
515 mL/min