Herunterladen Inhalt Inhalt Diese Seite drucken

NSK iCare Betriebsanleitung Seite 20

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
Guidance and manufacturer's declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an
environment.
Immunity test
IEC61326-1 Test Level
Electrostatic discharge (ESD)
±(2) 4kV Contact
EN/IEC61000-4-2
±(2, 4) 8kV Air
Electrical fast transient/burst
±2kV for power supply lines
EN/IEC61000-4-4
Surge
±1kV line(s) to line(s)
EN/IEC61000-4-5
±2kV line(s) to earth
Voltage dips, short
0% Ut
interruptions and voltage
(100% dip in Ut)
variations on power supply
for a cycles
40% Ut
input lines
EN/IEC61000-4-11
(60% dip in Ut)
for 10/12 a cycles*
70% Ut
(30% dip in Ut)
for 25/30 a cycles*
0% Ut
(100% dip in Ut)
for 250/300 a cycles*
Power frequency (50/60Hz)
30A/m
magnetic field
EN/IEC61000-4-8
NOTE: 'Ut' is the AC mains voltage prior to application of the test level.
*: eg: "25/30 a cycles" means "25 cycles for a 50Hz test", or "30 cycles for a 60Hz test".
Guidance and manufacturer's declaration - Electromagnetic Immunity
The product is intended for use in the electromagnetic environment specified below. The customer or the user of the product should assure that it is used in such an
environment.
Immunity test
IEC61326-1 test level
Conducted RF
3Vrms
EN/IEC61000-4-6
150kHz to 80MHz
Radiated RF
10V/m
EN/IEC61000-4-3
80MHz to 1.0GHz
3V/m
1.4GHz to 2.0GHz
1V/m
2.0GHz to 2.7GHz
NOTE1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should
be considered. If the measured field strength in the location in which the product is used exceeds the applicable RF compliance level stated above, the product should be
observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the product.
b: Over the frequency range 150kHz to 80MHz, the field strength should be less than 3V/m.
Compliance level
±(2) 4kV Contact
±(2, 4) 8kV Air
±2kV for power supply lines
±1kV line(s) to line(s)
±2kV line(s) to earth
0% Ut
(100% dip in Ut)
for a cycles
40% Ut
(60% dip in Ut)
for 10/12 a cycles*
70% Ut
(30% dip in Ut)
for 25/30 a cycles*
0% Ut
(100% dip in Ut)
for 250/300 a cycles*
30A/m
Compliance level
3V
10V/m
80MHz to 1.0GHz
3V/m
1.4GHz to 2.0GHz
1V/m
2.0GHz to 2.7GHz
19
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the product requires
continued operation during power mains interruptions, it is
recommended that the product be powered from an
uninterruptible power supply or a battery.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the product, including cables,
than the recommended separation distance calculated from
the equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2 P
d = 0.35 P 80MHz to 800MHz
d = 0.70 P 800MHz to 1.0GHz
d = 2.3 P 1.4GHz to 2.0GHz
d = 7.0 P 2.0GHz to 2.5GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer, and (d ) is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters as determined by
an electromagnetic site survey
(a)
should be less than the
compliance level in each frequency range
Interference may occur in the vicinity of equipment marked
with the following symbol:
(b)
.

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis

Fehlerbehebung

loading

Inhaltsverzeichnis