NSK iCare Betriebsanleitung Seite 11

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
4) When the Start Button is pressed lubrication starts. The Joint LED blinks during lubrication.
5) When lubrication is completed, a beep sounds and the Joint LED turns off.
• When Joint LED is off, lubrication for the joint will not start.
CAUTION
8-3 Removing Handpieces
• If debris is contained in the oil expelling from a handpiece even after lubrication, lubricate the handpiece again.
CAUTION
Removing Air Turbine
For Midwest 4 Hole Air Turbine (According to ISO 9168)
1) Open the door after the Joint LED has turned OFF and lubrication has completely finished.
2) Loosen the nut and pull out the handpiece straight from the Air Turbine Joint (Fig. 17).
For other types of air turbine handpieces
1) Open the door after the Joint LED has turned OFF and lubrication has completely finished.
2) Pull out the Air Turbine straight from the adaptor, or pull out the Air Turbine while pushing up the connector ring of the adaptor (Fig. 18).
Removing E-type handpieces (According to ISO 3964 type)
1) Open the door after the Joint LED has turned OFF and lubrication has completely finished.
2) Pull out the handpiece straight while pressing the Release Button on the E-type Handpiece
Joint (Fig. 19).
• With a cloth, wipe off any Oil collected at the bottom of the chamber and the
door. The collected oil could leak from the bottom of the door. Dispose of the
CAUTION
cloth as medical waste according to the laws and regulations in your country.
Nut
*C3 Type
10
Loosen
Fig. 17
*C2 Type
Release Button
*C2 Type
Adaptor
Fig. 18
Fig. 19