Herunterladen Inhalt Inhalt Diese Seite drucken

NSK S-Max M205 Benutzerhandbuch Seite 6

Luftmotoren
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9
Lubrication Oil
1) Remove the motor from the hose.
2) Apply 1-2 drops of lubrication oil into the Drive Air Port shown in the Fig. 7.
3) Connect the motor to the hose. Operate the motor for about 10 seconds.
7-3 Sterilization
Sterilize the product by autoclave sterilization. Remove the handpiece after each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
3) The motor should remain in the autoclave pouch until required for use.
8
CAUTION
· Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent possible discoloration and damage to the
product from chemical residue on other instruments.
· Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight. The air should be free from dust, salt and
sulphur.
· Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the product.
· If the sterilizer chamber temperature may exceed 135°C during the dry cycle then delete the dry cycle.
· Autoclave sterilization is recommended for the product. The validity of other sterilization methods is not confirmed.
· Do not touch the product immediately after autoclaving as it will be very hot and must remain in a sterile condition.
NOTICE
NSK recommends Class B sterilizers as stated in EN13060.
9

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis