1. Select SET-UP and press SET.
2. Select DATA and press SET.
3. Select DURATION and press SET.
4. Select the desired time for the data-logging and press SET.
5. Select LOG and press SET.
6. Decide whether or not all data (voltage, current, power, time) during
the next experiments should be logged and press SET.
7. Select SET-UP and press SET to return.
Starting a new experiment will erase the previous data in memory.
Stored data can be transferred to a computer via the RS232 output
using optionally available software.
At any time, you can return in the stand-by mode by holding SET
pressed during 2 seconds.
GROUND LEAKAGE

SHORT-CIRCUIT

OVERLOAD
CURRENT<1mA
CALL SERVICE
You could also try a general RESET: switch the instrument on
while holding SET pressed. Attention! All stored programs and data will be
erased!
EV2xx
EV2xx èwww.consort.be
www.consort.be
= A dangerous ground leakage is present (check
thoroughly the complete assembly). Your power
supply is NOT defective! See Frequently Asked
Questions at the end of this manual.
= Connected tank is short-circuited or has a very
low resistance (check thoroughly the complete
assembly). Your power supply is NOT defective!
See Frequently Asked Questions at the end of
this manual.
= Connected tank has a too low resistance (check
specifications). Your power supply is NOT
defective! See Frequently Asked Questions at
the end of this manual.
= The output current is lower than 1 mA (switch
the alarm off in the SET-UP menu). Your power
supply is NOT defective! See Frequently Asked
Questions at the end of this manual.
= Failure in the electronics. See Technical
Assistance Form at the end of this manual.

Data-logging

Data-logging
----
STAND-BY
SET-UP
----
SET-UP
DATA
-----
DATA
DURATION
LOG DATA
FOR:
-----
DATA
LOG
LOG
DATA:
-----
DATA
SET-UP
***
ERROR
GROUND LEAKAGE
***
ERROR
SHORT-CIRCUIT
***
ERROR
OVERLOAD
***
ERROR
CURRENT<1mA
***
ERROR
CALL SERVICE
----
----
-----
10h
-----
YES
-----

Alarms

***
***
***
***
***
10 /30
10
/30

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis