• When attaching the Nozzle, lightly thread the Nozzle screw into the Handpiece by hand
at first and then securely tighten it with the supplied Ring Wrench.
CAUTION
• After Ultrasonic Cleaning, blow remaining water from the inside with a syringe and
sufficiently dry the parts before use.
8-2 Sterilization
Sterilize the product by autoclave sterilization. Remove the Nozzle Tip and dispose of powder after each
patient and sterilize as noted below.
All parts of this product can be autoclaved.
Autoclave Procedure
1) Disassemble the product and clean it. (See 7-1 Discard of Perio Mate Powder - 7-4 Lubrication.)
2) Insert into an autoclave pouch. Seal the pouch.
3) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
4) Keep the product in the autoclave pouch to maintain sterility until it is used. To use, perform
reassembly in reverse order to disassembly.
• The Nozzle Tip is a single use product. Do not reuse or re-sterilize it since it may be
damaged or contaminated.
CAUTION
• Before use, thoroughly dry all parts.
• Remove the Powder Case Cap from the Powder Case, and then sterilize by autoclave.
Sterilization with the Powder Case Cap attached may cause breakage of the Powder
Case Cap.
• Do not autoclave the product with other instruments even when it is in a pouch. This is
to prevent possible discoloration and damage to the product from chemical residue from
other instruments.
• If the sterilizer chamber temperature may exceed 135°C during the dry cycle then
delete the dry cycle. Even in this case, sufficient drying of the product is required.
• Do not wipe with, or clean, or immerse in, high acid water or sterilizing solutions.
• Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation,
and sunlight. The air should be free from dust, salt and sulphur.
• Do not heat or cool the product too quickly. Rapid change in temperature could cause
damage to the product.
• The product can be sterilized only by autoclave sterilization.
• Do not touch the product immediately after autoclaving as it will be very hot and must
remain in a sterile condition.
NOTICE
• NSK recommends Class B sterilizers as stated in EN13060.
15