Herunterladen Inhalt Inhalt Diese Seite drucken

Maintenance; Cleaning - Tecno-gaz MASTER FLUX plus Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 82
ENGLISH
8
CHANGING THE CYLINDERS
If cylinder pressure is very low ~10÷20 bar, proceed with changing or re-charging the
cylinders.
Slowly close the valves (turn counter-clockwise).
Discharge the system by opening it from the "FLOW" and '%' knob, then close it back up.
Take the pressure reducers out, keeping them in a covered and clean location. (See
installation)
Re-charging, checking and periodic inspection of the cylinders must be carried out by
specialised companies in compliance with regulations in force in the country of use.
8.1

Cleaning

Use detergents that are not aggressive to plastic
Clean the surfaces of the device with a dry cloth.
Before and after use clean, wipe or sterilise the parts that come into contact with the
patient or his/her breath.
Cleaning table for MASTER FLUX parts
FIG/POS MATERIAL
3 / F
LATEX
3 / 2
VARIOUS
NON-TOXIC
11 / 29
PVC
THE STANDARD KIT IS DEPENDING ON THE MODEL
22

MAINTENANCE

CODE
DESCRIPTION
3MEDN0001
RUBBER BAG
SMFA506
FLOWMETER BOX
CM88051
SPIRAL TUBE
STERILISATION
STEAM
USING ETHYLENE
STERILISATION
NO
NO
NO
DISINFECTION
USING
QUATERNARY
OXIDE
SALTS
OK
OK
NO
NO
OK
OK

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Tecno-gaz MASTER FLUX plus

Inhaltsverzeichnis