Cleaning, care, and maintenance
The glass-ceramics material is largely chemically resistant, the heating zone remains plane at all times, is
free of pores and therefore easy to clean. If the heating zone is slightly dirty, you should at best clean it
simply using warm water and a few drops of a household washing-up liquid as long as the heating zone is
lukewarm or cold.
To remove crust, lime, and water marks or stains with a metallic shine you should use a household deter-
gent for ovens with glass-ceramics boiling zones. Please use a wet cloth or sponge to remove the deter-
gent residues as well, since some of the commercially available detergents adopt an etching effect at
higher temperatures. If soiling has already turned into a crust, you should use a razor-blade scraper. Plas-
tic objects and aluminium foil as well as sugar-containing substances reaching the hot surface are to
be lifted off immediately using the razor-blade scraper.
6

Malfunction, possible defects

Malfunction / defect
The Laboratory hot plates
does not switch on;
heating zone does not heat up
No display
Impossibility of switching the
Laboratory hot plates off
In the case of the SLK 6 no
display of the temperature,
but
Never use abrasive cleaners! In most cases it will suffice after each use to
clean the plates using a moist cloth and some household washing-up liquid.
Subsequently rub dry.
Crusts and food boiled over should at best be first soaked with a wet cloth.
Subsequently, use a glass scraper to remove the residues.
Note:
Sugar and melted plastics should be removed immediately when the heating
zone is still hot.
Cause
Sensor field not clean
No mains voltage
Defective heating zone;
defective electronics
7-segments display or
corresponding electronics
defective
Electronics defective
Broken temperature sensor;
temperature sensor
is missing
20
Moist cleaning of sensor field, rub dry
Check mains cable including its connectors;
check fuse of the mains-socket circuit,
replace if necessary
Separate Laboratory hot plates from current
supply: separate from mains;
send Laboratory hot plates in for repair
Caution! Danger of burning!
Separate Laboratory hot plates from mains;
send Laboratory hot plates in for repair
Separate laboratory hot plates from mains;
send Laboratory hot plates in for repair
Check connection of temperature sensor;
if necessary, replace temperature sensor
Remedy