Open door. All movements are in the zero position,
i.e. the chair has been lifted max up and in, rotated
max in and moved max back.
Open door. Chair max down.
Messages on error, numbered 1-18.
Low battery voltage.
Cable brake in the manual control.
The back support is leaning too far back.
1 Lengthwise adjustment motor overloaded.
2 Fault on pulse sensor for longitudinal adjustment
1 Rotation motor overloaded.
2 Fault on pulse sensor at rotation motor.
26
motor.
Right – drive forward.
Down – rotate and lower.
Up – raise and rotate in.
Remedial actions.
Charge the battery.
Change cable
Raise the back support.
1 Check that nothing is obstructing the movement.
2 Drive in the emergency-mode operating position,
then contact the installer.
1 Check that nothing is obstructing the movement.
2 Drive in the emergency-mode operating position,
then contact the installer.