3. Error Messages

The instrument displays error messages when an error occurs.
Check and respond to them correctly.
Error
Messages
Insufficient light for
measurement.
Flashing
Pulse signal required
for measurement is
not being received.
Flashing
The pulse is too strong
for correct measure-
ment.
Flashing
*1 Contact the nearest authorized service facility.
E4
Cause
Solution
Check that the finger holder
1
is properly attached to the
patient.
If the part being measured is
2
thick, insufficient light will be
transmitted. Use another, thin-
ner finger.
Clean the light-source and
3
sensor sections.
If this message persists, the
4
instrument has likely failed. *1
Properly insert a fingertip into
1
the finger holder.
This is likely caused by insuf-
2
ficient blood circulation to the
fingertip. Rub or warm the
fingertip or use another finger.
Properly insert a fingertip into
1
the finger holder.
This is likely caused by exces-
2
sive blood circulation to the
fingertip. Take a rest for a while
and retry again or use another
finger.
If this message persists, the
3
instrument has likely failed. *1