5. Signal Adjustment

Lissajous
Oscilloscope
CH2
CH1
Move the detector head in the azimuth direction shown in the figure, and then adjust so that the Lissajous is
within the range of the specified values over the entire length.
<Specified values>
A/B signal output type ...... Output format A : Amplitude 0.9 Vp-p to 1.8 Vp-p
Analog output type ........... Output format H : Amplitude 0.3 Vp-p to 0.6 Vp-p
Azimuth direction
• For output format A, the center of the Lissajous' figure is 0 V. For output format F, G, and H, the center of
the Lissajous' figure is 2.5 V.
• Turn off the interface unit before connecting or disconnecting the oscilloscope.
BL55-NE
CH1
Output format A
CH2
TP301
( SIN )
TP351
( COS )
Probe
TP61
(AGND)
TP301
(SIN)
TP62
TP201 (A)
(DGND)
TP251
TP351
(COS)
(B)
Interface unit
Output format F : Amplitude 0.7 Vp-p to 1.3 Vp-p
Output format G : Amplitude 0.8 Vp-p to 1.2 Vp-p
TP61
( AGND )
Signal adjustment
section-detailed view
Output format F, G, H
CH1
TP3
CH2
TP2
TP1
Azimuth direction
TP6
TP8
TP5
TP4
TP7
5
(E)