Herunterladen Inhalt Inhalt Diese Seite drucken

OHAUS DEFENDER 3000 Bedienungsanleitung Seite 10

Bodenwaage
Vorschau ausblenden Andere Handbücher für DEFENDER 3000:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
EN-10
DEFENDER 3000 SERIES FLOOR SCALE,
Installation of Optional Foot Plates and Ramps
Plates are required for certifiable installations. Installation of each Ramp also requires the installation of one set of
Foot Plates. Foot Plates are available separately and are included with the Ramp accessory kit. Refer to Figure 3
for mounting of the Foot Plates and Ramps. Installation of Floor Plates and Ramps must be done before final
placement of the Floor Scale.
Notes: 1. Expanding bolts require 12.5mm diameter holes drilled 60mm deep into the floor surface.
2. Foot Plate center to center dimensions W and L are shown in Table 6.
3. The Floor Scale must be positioned so the load cell feet rest within the large holes of the Foot Plates.
4. Ensure that a free air gap exists between the sides of the Floor Scale and the edges of the Ramps.
TABLE 6. DIMENSIONS FOR FLOOR PLATE AND RAMP INSTALLATION
MODELS
DF1500BS
DF1500BR
DF1500BL
DF3000BL
DF1500BX
DF3000BX
DF3000BV
DF1500BRL
DF1500BLX
DF3000BLX
Figure 3. Optional Foot Plate and Ramp Installation
RAMP
A
800
1000
1200
1500
2000
1200
1500
DEFENDER 3000 Series Floor Scale
DIMENSIONS (mm)
FOOT PLATE CENTERS
B
W
800
700
1000
900
1200
1100
1500
1400
2000
1900
1000
900
1200
1100
L
700
900
1100
1400
1900
1100
1400

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis