2. Temperature adjustment
Wird bei der Temperaturkalibrierung die zulässige Abweichung über-/unterschritten
kann die Temperatur des Instruments wie nachfolgend beschrieben justiert werden.
If the admissible deviation is exceeded / not reached during temperature calibration the
temperature of the instrument can be adjusted as described below.
1. Switch off instrument.
2. Press
„8.1x" will be displayed. If not, separate device from power supply and restart
step two.
3. Adjust temperature using arrow keys
ton the temperature will be increased or reduced by 1°C.
4. Confirm input by
For further information on how to operate the moisture meter please refer to the
operating instructions enclosed with each model.
DAB_A01-BA-def-1510
, quickly release, then press
.
and
at the same time.
. Each time you press the but-