Technical information
E
Failure
Possible cause
encountered
Display 0
Start/Stop button accidentally
0
pressed while battery was
being changed.
Measure-
Unit detects a fault or that
ment was
diastolic value is too low.
interrupted
Movement during
and cuff first
measurement.
deflated and
then inflated
No display
Batteries inserted incorrectly. Check position of batteries.
after unit is
Batteries flat.
switched on
Battery contacts dirty.
No display
No measured values stored. Take a new measurement.
after Memory
button is
pressed
Meas-
Batteries flat.
urement
interrupted
Failure
Possible cause
encountered
EN-54
Corrective action
Switch unit off and then on
again with Start/Stop button, if
necessary.
Unit shows no fault. Repeat
measurement.
If necessary, interrupt meas-
urement, measure again after
a break of 5 minutes.
Change batteries.
Clean battery contacts.
Change batteries.
Corrective action
Failure
Possible cause
encountered
Measured
Incorrect measurement
values
position.
extremely
high or low
Date/time
Function accidentally
not shown in
deactivated. Function was
display
not reactivated after battery
replacement.

2. Customer service

The unit may only be repaired by the manufacturer or by an expressly
authorised organisation. Please contact:
UEBE Medical GmbH
Zum Ottersberg 9
97877 Wertheim, Germany
Tel.:
+49 (0) 9342/924040
Fax:
+49 (0) 9342/924080
E-mail:
info@uebe.com
Internet:
www.uebe.com
Technical information
E
Corrective action
Repeat measurement,
ensuring correct body posture.
Do not move your arm and do
not talk.
Activate date and time as
described on Page EN-42.
EN-55