Test Frequency
(MHz)
Radiated RF
1720
IEC61000-4-3
1845
(Test specifications
for ENCLOSURE PORT
1970
IMMUNITY to RF wire-
less communications
equipment)
2450
5240
5240
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM
may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it
would be worst case.
Band a)
Service a)
(MHz)
1 700 – 1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
2 400 – 2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
5 100 – 5 800
WLAN 802.11
a/n
Modulation b)
Modulation b)
(W)
Pulse
2
modulation b)
217 Hz
Pulse
2
modulation b)
217 Hz
Pulse
0,2
modulation b)
217 Hz
74
Distance
IMMUNITY TEST LEVEL
(m)
(V/m)
0,3
28
0,3
28
0,3
9