When possible no abrupt angles or other disturbances shall be
mounted directly in front of the instrument.
All devices have metric inlet threads (G-thread = BSPP). The
mounting of NPT-fittings will destroy the inlet threads in the
instrument's body.
Before the first start-up it needs to be checked that the system
has absolutely no leakage. If toxic, explosive and/or corrosive
gases are to be used, ensure to purge the system enough with
a dry, inert gas for at least 30 minutes prior and after being
used. This is also required to remove these kinds of gases from
the system.
The instruments' preferred mounting position is horizontally.
Especially when installing bigger mass flow controllers in
different ways you should get in contact with your distributor or
M+W Instruments GmbH beforehand.
All instruments described in this manual carry the CE-mark.
Therefore they have to comply with the EMC requirements valid
for these instruments. However compliance with the EMC
requirements is not possible without the use of proper cables
and connector/gland assemblies. In case the instrument is
connected to other devices (e.g. power supply) ensure that the
operating voltage complies with the supply voltage given on the
name plate and that the function of the shielding is not affected.
M+W Instruments GmbH recommends using standard cables
following the CE-requirements and could supply them when
needed.
19/32