2.13 Notice of Operating

a. For Waveform Measurement :
The PeakTech
oscilloscope input impedance must be matched to 50Ω. Use the coaxial cable for
characteristic impedance 50Ω in connecting with PeakTech
Out and oscilloscope input terminal.
Minimizing the cable length and cable stray capacitance is very important for the best
performance.
Because the function generator output is a wideband signal, every connecting path
including the transmitter or receiver, must be impedance matched to 50Ω, in order to
avoid the reflection from load and the undesired testing results.
P W H : 1 . 0 0 0 0
5 0 0 . 0
Pulse Width Frequency Setting
D T Y : 5 0 . 0
5 0 0 . 0
Pulse Width Duty Setting
F R Q : 1 . 0 0 0 0
F M
O F F
FM On, Internal 400Hz Source
FM On, Internal 1000Hz Source
®
1007 / 1014 / 1016 Func Out output impedance is 50Ω, so the
k H z
u S
S B 5
%
u S
S B 6
k H z
S B 7
FM Off
®
1007 / 1014 / 1016 Func
60