2.13 Notice of Operating
a. For Waveform Measurement :
•
The PeakTech
oscilloscope input impedance must be matched to 50Ω. Use the coaxial cable for
characteristic impedance 50Ω in connecting with PeakTech
Out and oscilloscope input terminal.
•
Minimizing the cable length and cable stray capacitance is very important for the best
performance.
•
Because the function generator output is a wideband signal, every connecting path
including the transmitter or receiver, must be impedance matched to 50Ω, in order to
avoid the reflection from load and the undesired testing results.