Electromagnetic Compatibility Information
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic emissions
The ME 90 is intended for use in the electromagnetic environment specified below. The customer or the user
of the ME 90 should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions Complies
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer's declaration – electromagnetic immunity
The ME 90 is intended for use in the electromagnetic environment specified below. The customer or the user
of the ME 90 should assure that it is used in such an environment.
IMMUNITY test IEC 60601
test level
Electrostatic
±6 kV
discharge (ESD)
contact ±8
IEC 61000-4-2
kV air
Electrical fast
±2 kV for
transient/burst
power sup-
IEC 61000-4-4
ply lines
Surge
±1 kV line(s)
IEC 61000-4-5
to line(s)
±2 kV line(s)
to earth
Compliance
Group 1
Class B
Class A
Compliance
Electromagnetic environment – guidance
level
±6 kV contact
Floors should be wood, concrete or ceramic tile. If floors
±8 kV air
are covered with synthetic material, the relative humidity
should be at least 30 %.
±2 kV for power
Mains power quality should be that of a typical commer-
supply lines
cial or hospital environment.
±1 kV line(s) to
Mains power quality should be that of a typical commer-
line(s)
cial or hospital environment.
±2 kV line(s) to
earth
40
Electromagnetic environment – guidance
The ME 90 uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
The ME 90 is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.