Note regarding state of logger upon delivery:
Upon its delivery the logger is in a kind of 'sleeping state': the display does not show anything, the power consumption is at
its minimum.
The EASY
'wakes up' as soon as it is connected to an EASY
LOG
with a software has been established. The display jumps back and forth between the current measuring value and 'Stop' and
the logger is ready for operation.
Programming of the display range:
The display range, position of decimal point and the display unit can be set by the EBxKonfig software. Furthermore, it can
be used to read out sensor information
Adjustment of EASY
The assignment of the angle-type plug is designed for the most commonly used assignments of the respective input signals.
As this is not a standardised assignment, your transmitter assigment may not correspond to the EASY
How to change the assignment of the angle-type plug:
Dismantle the plug by pulling the adaptor inset out of
the case, using a screw-driver at the lateral groove.
Change the assignment according to the notes at the
respective input signal.
Latch coupling insert in cover. You have a choice
between 4 different orientations - each of them spaced
90°.
Please make sure to adjust the conneting cable, if the
assignment of the angle-type plug has been changed
for the types 4-20mA and 0-20mA.
Put on angle-type plug and connect plugs using the
long screw delivered. (Do not forget seals).
Input signal: 4-20mA
In the angle-type plug the male contacts
1, 3 and 4 are directly connected 1:1
with the socket. The EASY
is located between the male contact 2 (-)
and the jack 2 (+).
pin contact
1
2
3
4
_
Rs
100Ohm
+
1
2
3
4
jack
If the 'signal/GND'-line in your transmitter
is not assigned to contact 2, please do
not forget to adjust the EASY
type plug and the external angle-type
plug accordingly:
To do so open the EASY
plug and exchange the red wire of
contact 2 against the wire of the contact
representing the 'signal/GND' in your
transmitter. Then exchange and/or rewire
the two contacts in the angle-type plug of
your connecting cable.
Connection advice:
If more than one EASY
BUS
module (e.g. EASY
4 0 N S . . ., EASY
LOG
(unit type, serial number, address, etc.)
-connections:
LOG
Input signal: 0-20mA
In the angle-type plug the male contacts
1, 2 and 4 are directly connected 1:1
with the socket. The EASY
LOG (Rs=100Ohm)
is located between the male contact 3 (-)
and the jack 3 (+).
1
2
1
2
If the 'signal'-line in your transmitter is
not assigned to contact 3, please do not
forget to adjust the EASY
LOG
-angle-
plug and the external angle-type plug
accordingly:
To do so open the EASY
LOG
-angle-type
plug and exchange the black wire of
contact 3 against the wire of the contact
representing the 'signal' in your
transmitter. Then exchange and/or rewire
the two contacts in the angle-type plug of
your connecting cable.
-sensor modules are connected at the same EASY
4 0 I M P, EBN) has to be isolated from the others.
BU
Bus
-level converter (e.g. EBW1) and a communication link
internal assignment of angle-type plug:
table of connections angle-type and jack
wire
colour-
Pin no.:
blue
1
red
2
black
3
yellow
4
LOG (Rs=100Ohm)
pin contact
3
4
_
Rs
100Ohm
+
3
4
jack
LOG
-angle-type
LOG
-angle-type
and to program the alarm points of the logger.
Inputs
4-20mA
0-20mA
feeded through
feeded through
EASYLOG
feeded through
(Rs = 100 Ohm)
EASYLOG
feeded through
(Rs = 100 Ohm)
feeded through feeded through
Input signals: 0-2V, 0-10V
In the angle-type plug the 4 male
contacts are directly connected 1:1
with the socket. The EASY
connection is on contact 1
(signal+) and contact 2 (signal-).
If your transmitter assignments for
the 'signal+' and 'signal-' are
different, please do not forget to
adjust male contacts of your angle-
type plug accordingly:
To do so open the angle-type plug
and remove the red and the blue
wires entering the housing from the
coupling of the angle-type plug.
Connect wires with the respective
contacts, representing signal +
(blue wire) and signal - (red wire) at
your transmitter.
, the input signals of each EASY
BUS
E25.0.2X.6C-03
LOG
assignment.
:
0-2V, 0-10V
feeded through
(EASYLOG, signal +)
feeded through
(EASYLOG, signal -)
feeded through
feeded through
LOG
BUS
-sensor