Herunterladen Inhalt Inhalt Diese Seite drucken

Error Codes - Adam CBK 4 Bedienungsanleitung

Cbk-m series
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 53
EN
8.0

ERROR CODES

During the initial power-on testing or during operation, the scale may show an error message. The meaning
of the error messages is described below.
If an error message is shown, repeat the step that caused the message. If the error message is still shown
then contact your dealer for support.
ERROR
DESCRIPTION
CODE
Err 1
Time input Error
Err 2
Date input Error
Err 4
Initial Zero is greater than allowed
(4% of maximum capacity) when
power is turned on or when the
[Zero/Enter] key is pressed.
Err 6
A/D count is not correct when turning
the scale on.
Err 7
Percent input error
Err 8
High limit input error
Err 9
Low limit input error
FAIL H or
Calibration error
FAIL L
"Err 1"
Err 2"
NOTE: The
,
applications.
© Adam Equipment Company 2013
FAIL H or FAIL L
and
are not applicable for CBK-M Scales for use in approved
14
POSSIBLE CAUSES
Invalid time entry such as "268970" for the time
"H-m-S".
format
th
34
day of a month is an invalid entry.
Weight on the pan when turning the scale on.
Excessive weight on the pan when zeroing the
scale.
Platform is not installed.
Improper calibration of the scale.
Damaged load cell.
Damaged Electronics.
Load cell is damaged.
Electronics is damaged.
Percent function is entered with no reference
mass on the pan.
Low limit is set first, and then the high limit is set
lower than the low limit and high limit not equal
to zero.
High limit is set first, then the low limit is set
higher than the high limit and low limit not equal
to zero.
Improper calibration (should be within +10% of
the factory calibration). The old calibration data
will be retained until the calibration process is
complete.
3026610536RevA3-Jul13

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cbk 8hCbk 48Cbk 8Cbk 16Cbk 32

Inhaltsverzeichnis