4
not activated when lowering the table. If you do
not succeed in lowering the nursing table com-
pletely the first time, then perhaps try one more
time.
2. Once the nursing table has been completely
lowered, release the lowering button, and after
3-4 sec. press it in again until you hear a slight
click after approx. 10 sec.
3. Release the lowering button and then raise the
230V, 50/60 HZ
nursing table to its highest position. You will hear
3000 mm
another click.
4. Lower the nursing table a couple of centimetres.
5. Then raise it again to its highest position. You will
hear another click. The nursing table is now reset.
The nursing table should also be reset if there has
been an interruption to the power supply.
Service:
It is recommended that the nursing table is serviced
every 12 months The procedure is as follows:
•
Remove the plug to the safety tray.
•
Remove the cover plate.
•
Check the moving parts for wear and check
whether all cables are intact.
•
Check that the wall support is properly secured to
the wall.
•
Remove the safety tray and clean thoroughly.
•
Check that the wire baskets run freely in and out
on their rails.
•
Check that the feed and waste pipes are tight and
intact.
•
If necessary, dismantle the water trap and clean
it.
•
Check the mixer tap.
•
Replace the safety tray under the table.
•
Replace the cover plate
•
Insert the plug for the safety tray into the control
box.