EMC Tables
Guidance and MANUFACTURER'S declaration – ELECTROMAGNETIC EMISSIONS
– FOR ALL THE EQUIPMENT AND THE SYSTEMS
I-PRESSS 4 is intended for use in the electromagnetic environment specified below. The
customer or the user of the I-PRESS 4 should assure that it is used in such an
environment.
Emission test
RF emissions
Cispr 11
RF emissions
Cispr 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and MANUFACTURER'S declaration – ELECTROMAGNETIC IMMUNITY
– FOR ALL THE EQUIPMENT AND THE SYSTEMS
I-PRESS 4 is intended for use in the electromagnetic environment specified below. The
customer or the user of I-PRESS 4 should assure that is used in such environment.
Immunity test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical
fast
transient/burst
IEC 61000-4-4
IACER SRL
Compliance
Group 1
Class B
Class A
Complies
Complies
Test
level
EN
60601-1-2
± 6kV contact
± 8kV air
± 2kV for power
supply lines
Page 20
Electromagnetic environment –
guidance
I-PRESS 4 uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment
I-PRESS 4 is suitable for use in all
establishments,
establishments
and
connected to the public low-voltage
power supply
network that supplies
buildings used for domestic purposes.
The I-PRESS 4 is suitable for use in all
establishments, other than domestic
establishments
and
connected to the public low voltage
power supply network that supplies
buildings used for domestic purposes
Compliance level
± 6kV contact
± 8kV air
± 2kV per power
supply lines
including
domestic
those
directly
those
directly
Electromagnetic
environment
–
guidance
Floors
should
be
wood, concrete or
ceramic tile. If floor
are
covered
with
synthetic
material,
the relative humidity
should be at least
30%
Mains power quality
should be at that of
a typical commercial
or
hospital
environment.
MNPG220-00 25/03/16