Ambient conditions

-
The ZA-TEC ZA 1-3 is intended for measurements under dry ambient condi-
tions,
-
Maximum barometric height for measurements: 2000 m,
-
Overvoltage category / installation category: IEC 60664-1/ IEC 61010-1 →
600 V category III; 1000 V category II,
-
Contamination class: 2,
-
Protection category: IP 30 DIN VDE 0470-1 IEC/EN 60529
IP 30 means: Protection against access to dangerous parts and protection
against solid impurities of a diameter > 2.5 mm, (3 – first index).
No protection against water, (0 – second index).
-
Operating temperature and relative air humidity:
For operating temperatures from 0 °C to 30 °C: relative air humidity less
than 80 %,
For operating temperatures from 31 °C to 40 °C: relative air humidity less
than 75 %,
For operating temperatures from 41 °C to 50 °C: relative air humidity less
than 45 %,
-
Storage temperature: The ZA-TEC ZA 1-3 can be stored at temperatures
between - 15 °C and + 60 °C (air humidity of 0 to 80 %). During storage, the
battery should be removed.
7.
Electrical specifications
Note: The measuring accuracy is specified as the sum of:
-
a relative part of the measured value and
-
a number of digits (i.e. counting steps of the last digit).
This measuring accuracy applies for temperatures from 18 °C to 28 °C and a
relative air humidity less than 80 %.
7.1 DC voltage ranges
The input resistance is 10 MΩ.
Measuring
Resolution
range
*3
200 mV
2 V
20 V
200 V
1000 V
7.2 AC voltage ranges
The input resistance is 10 MΩ in parallel 100 pF.
Measuring
Resolution
range
*3
200 mV
100 µV
2 V
1 mV
20 V
10 mV
200 V
100 mV
750 V
The measured value of the ZA-TEC ZA 1-3 is obtained by mean value rectifica-
tion and is displayed as r.m.s. value.
The measuring accuracy is specified for a sinusoidal curve. In case of non-
*1
sinusoidal curves, the accuracy of the displayed value decreases. Thus, an
additional error results for the following crest factors:
Crest factor from 1.4 to 3.0 additional error ± 1.5 %
Crest factor from 3.0 to 4.0 additional error ± 3 %
Applies to sinusoidal curves of 50 Hz/ 60 Hz
*2
For automatic range selection (AUTO), the change-over point might be al-
*3
ready at a value of 1400!
7.3 DC current ranges
Overload protection:
-
10 A (600 V) fuse, quick-acting, 50 kA, at the 10 A input,
11/ 2016
100 µV
± (0.5 % of the measured value + 2 digits)
1 mV
± (0.5 % of the measured value + 2 digits)
10 mV
± (0.5 % of the measured value + 2 digits)
100 mV
± (0.5 % of the measured value + 2 digits)
1 V
± (0.5 % of the measured value + 2 digits)
Measuring accuracy *
frequency range 50 Hz - 300 Hz
± (2.0 % of the measured value + 5 digits) *
± (1.5 % of the measured value + 5 digits)
frequency range 50 Hz - 500 Hz
± (1.5 % of the measured value + 5 digits)
± (1.5 % of the measured value + 5 digits)
1 V
± (1.5 % of the measured value + 5 digits)
ZA-TEC ZA 1-3
Measuring accuracy
1
within the
Overload
protection
1000 V
DC
1000 V
DC
1000 V
DC
1000 V
DC
1000 V
DC
Overload
protection
2
750 V
eff
750 V
eff
750 V
eff
750 V
eff
750 V
eff
14