3. Moduses and Functions
GB
34
3.1.2.3 TC-Probe
This menu is used for configuring the thermocouple calibration.
Fig. 18: TC-probe setup
menu
There are two possibilities:
binding posts 2 and 3
-> input 2
Thermo-socket TC
-> input 3
Fig. 19: Connectors of the
instrument
Binding Posts
The binding posts are suitable for connecting blank wire ends,
spade connectors and 4 mm banana plugs.
Thermo-Socket
The thermo-socket can be used to connect standard thermo-
connectors and miniature thermo-connectors.
Always observe the correct polarity when connecting thermo-
couples. If thermocouples need to be extended, the thermo-cable
or extension cable of the respective thermocouple type must be
used.
To calibrate thermocouples, the
respective type of
thermocouple is selected with
the cursor keys and confirmed
with the SELECT key.
Subsequently, the input
sockets must be defined.