Shutdown Mode Settings
The shutdown mode is used to determine the automatic test end point of the moisture analyzer and display the
results. The moisture analyzer has two shutdown modes to set: timed shutdown and automatic shutdown.
»
Timed shutdown: manually set the test time.
Press the time button and the time symbol flashes. The hydrometer displays the default time of the system or
the time set last time (minutes).
Reduce the temperature setting for 1 minute every time you press Tare, increase the temperature setting for 1
minute every time you press Start. Press [Time] button to confirm the save.
»
Automatic shutdown: automatically set the test time.
Press [Menu] to display AUTO.
Press the "Start" button to confirm that the setting is successful when the letter "OK" appears in the lower left
corner of the display, and the instrument will return to the weighing state.
(If you want to change the automatic mode to manual mode, press [Menu] button the instrument displays
"AUTO", then press the [Tare] button to find "CLOCK" and press the [Start] button to confirm, the setting is
successful when "OK" appears in the lower left corner of the display, and the instrument will return to weighing
mode.)
Loss Rate Settings
»
A1 (5mg/20s): Rapid trend detection.
»
A2 (5mg/50s): Quick drying.
»
A3 (5mg/60s): Standard setting for most samples.
»
A4 (5mg/90s): For medium-dry samples.
»
Press [Menu] button to enter the option and press the [Tare] button to select.
»
Press the [Start] button to confirm, and the instrument will display the graphic symbol of automatic shutdown
and return to the normal weighing state.
9. WARNING
The following substances may pose a risk of fire, explosion, damage, or injury. For substances with potential
safety hazards, please carefully analyze the possible hazardous consequences. In this case, the instrument
should be watched by a dedicated person and wear corresponding protective devices.
1. Volatile substances
For volatile substances, it is recommended to use a rapid heating method to limit the loss of water before the
sample evaporates. The test should be conducted in a safe and dry environment with a sufficiently low
temperature to prevent explosion or fire. Please use 1 g or less of sample for testing.
2. Toxic substances
Contains toxic or corrosive substances and should be dried in ventilated shop windows.
3. Corrosive substances
Heating substances that produce corrosive gases, such as acidic substances, should be tested with as few
samples as possible. Because of its production.