• Connect the short silicon tube, with filter antibacterial, to the suction connector .
The other tube, with one end connected to the filter must be connected with the
other end to jar's lid connector where has been fixed the red float. When the 90% of
the volume of the jar is reached there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid liquid penetration inside the
device. The device must be used on a plan of horizontal operation.
• Connect the long silicon tube to the other jar's lid connector
• Connect the other end of the long silicon tube to the probe plastic connector then connect
the suction probe to it.
• Connect the power cord to the device then connect the plug to the electrical mains
supply.
• Push switch on position I to start suction
• Once finished push switch on 0 position and unplug.
• Unscrew the jar's lid and fill the jar 1/3 full or ordinary water (this for an easy
cleaning operations and an rapid reaching of the functionally vacuum) then rescrew
the lid on the jar correctly.
• To extract the accessories and start with cleaning.
The dual use version is available on request.
Continuous use can be enabled by activating the illuminated switch.
Alternatively, the pedal control can be used for aspirating on command, while leaving
the operator's hands free. If you're using the pedal version, insert the plug of the pedal
power cord into the appropriate socket located next to the equipment, push the second
switch on the position II, so start and stop the suction operation by pressing relasind
the pedal.
Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted into the "VACUUM".
COMPLYING WITH THE NEW EUROPEAN RULES, MORETTI INDICATES THE IMPORTANT POINTS TO PROTECT INSTRUMENT AND OPERATORS
HYGIENE. THESE RULES MUST BE RESPECTED IN ORDER TO GUARANTEE HYGIENE AND SAFETY TO ALL THE PEOPLE OPERATING WITH THE
INSTRUMENT TO OBTAIN QUALITY AND WELL BEING.
Every returned instrument will be hygienically checked before repairing. If MORETTI finds instrument not suitable for repairing due to clear signs of
internal or external contamination, the same will be returned to customer with specification of NOT REPAIRED INSTRUMENT, accompanied by an
explanation letter. MORETTI will decide if contamination is due to bad functioning or misuse. If contamination is due to bad functioning, MORETTI
will substitute the instrument, only if a SALE RECEIPT and STAMPED GUARANTEE accompany the same.
MORETTI is not responsable for contaminated accessories, they will be substitute at customer's expenses.
For this reson it is COMPULSORY to carefully disinfect the external part of the instrument and accessories with a cloth soaked in methylated spirits
or hypochlorite-based solutions. Put the instrument and accessories in a bag with indication of disinfecting. We also request to specify the kind of
fault, in order to speed up repairing procedures. To this end, please read the instructions carefully in order to avoid damaging the equipment
through improper use. Always specify the fault encountered so that MORETTI can establish whether it falls into the category of the faults covered
by the guarantee.
INSTRUCTION FOR USE
A wrong connection causes immediate destruction in case of contact with sucked liquids.
RULES FOR RETURNING AND REPAIRING
13