Also distinguish between:
•
•
NOTICE
Incorrect position values from relocating the machine
zero or changing the encoder type.
Position correction values are valid as long as the location of
the machine zero and the chosen encoder type are unchanged.
If you, after entering position correction values,
•
•
•
•
you must delete all previously stored position correction values
and enter new position correction values.
154
Reference switch: In your measuring stage from Märzhäu-
ser Wetzlar, integrated references switches generate a con-
trol signal when they are switched, and thus the accur-ate
reproducible machine zero for the position correction.
Reference mark: Measuring systems from other manu-
facturers can be equipped with one or several reference
marks, which deliver a separate control signal when they are
approached. This control signal can also be used for setting
the machine zero for the position correction.
relocate the machine zero,
approach another reference mark to set the machine zero,
change the function for determining the machine zero or
change the encoder type,