Connection cable mount
The probe has an M8 connection according to IO-Link port type A. The pin assignment
is as follows.
4
2
3
1
Pin number
Description
1
V+ (24V)
2
not used
3
V- (GND)
4
IO-Link CI/CO
According to the IO-Link specification Port A, a 4-pin M8 A coded (usually on M12)
sensor cable must be used to connect the probe. This may be a maximum of 10 m long.
A shielding of the wires is not specified. The sensor is thus connected to a suitable
IO-Link master.
Application
Pneumatic advanced transducers
The pneumatically pushed transducer have the following specified maximal pressure
P models - 1.5 bar
L models - 4.5 bar
Application of pneumatic air to push forward the tip is allowed only when contacting a
workpiece. Otherwise the stroke limitation of the linear ball bearing can permanently
be destroyed!
IO-Link Specification und IODD
The sensor is designed and compliant with the IO-Link specification 1.1. All sensors
have the DeviceID 1000, which means that they all share the same protocol functions.
The ProductID according to IO-Link corresponds to the article number. An IODD file is
available for all probes. This can be obtained from www.peterhirt.ch on the correspon-
ding product page.
IO-Link process data
The probes only have process in data. These are 32 bit wide and correspond to the
position of the measuring pin in nanometers, signed.
Subject to change without notice. Errors and omissions excepted