Malfunctions
 Do NOT attempt to open the chamber door during a cycle or in the event of a power
failure or system malfunction (see section B5.1).
 All repairs must be carried out by trained personnel ONLY. Do NOT attempt to repair or
modify any part of the device. The trophon EPR contains no end user serviceable parts.
Transporting the Device
 The device weighs approximately 17kg (38lb). Use safe lifting techniques as per your
Occupational Health and Safety Lifting Guidelines for your institution.
 Do not move, relocate or transport the device if hydrogen peroxide is present; purge
the device before moving or relocation (See section B3).
Electrical Device
 Equipment must be connected to an earthed power outlet. Ensure power cable
supplied with the device is used.
 Spilled fluid can result in electrical shock. Do not allow any fluid to spill onto or around
the device or immerse any parts of the trophon EPR in liquid.
 Connect the device only to an electrical source with the proper voltage and
frequency as specified in Appendix 1. If incorrect voltage is used, the device may be
damaged when it is switched on.
 Attempting to open any component of the device to gain access to the internal
mechanics may result in electrical shock.
trophon Chemical Indicator
 ONLY to be used with the trophon EPR. Do NOT use with non-approved devices.
 Use one Chemical Indicator with every trophon EPR cycle.
 Do NOT use damaged and/or out-of-date indicators.
 Ensure indicator is red and not exposed before use.
 More information can be found in the Chemical Indicator Instructions for Use, provided
with the Chemical Indicator.
trophon NanoNebulant
 Read the Instructions for Use (IFU) leaflet enclosed with the trophon NanoNebulant and
the Material Safety Data Sheet (MSDS) enclosed with the device.
 Always wear disposable gloves when handling disinfectant cartridges and running the
trophon EPR device.
 Ensure appropriate personal protective equipment (PPE) is worn during spills
(according to the Occupational Health and Safety Guidelines for your institution and
NanoNebulant MSDS).
®
trophon
EPR User Manual – N00020
All manuals and user guides at all-guides.com
L00034 1.0 03/2014