Introduction
The PACE5000 single-channel and PACE6000 single/dual-channel, Pressure Automated
Calibration Equipment measures and controls pneumatic pressures and displays, on a touch-
screen, the pressure measurement and controller status. The touch-screen enables selections
and settings in both measure and control modes. The instrument can be operated remotely
through communication interfaces.
For the full specification and user manual, refer to Druck website:
Safety
WARNING
other strong oxidizing agents.
This product contains materials or fluids that may degrade or combust in the
presence of strong oxidizing agents.
Do not use this device for any other purpose than that stated.
Do not apply pressures above the Maximum Working Pressure (MWP) stated on the rear
panel (Refer to Figure A1, item 5).
Do not apply electrical power above the maximum values stated on the rear panel (Refer to
Figure A1, item 2).
Refer to K0443 PACE User Manual for general pressure equipment requirements.
Maximum Working Pressure (MWP): Stated on rear panel
Pressure Media: Clean, dry, nitrogen or air compatible with stainless steel, acrylic and nitrile.
Pressure Connections: ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202) (filtered).
Note: Refer to Data Sheet for the complete range of adaptors.
Do not use with media that has an oxygen concentration > 21 % or
https://druck.com
English–PACE5000/6000 Safety Instructions | 1