Herunterladen Diese Seite drucken

Baker Hughes Druck SENSYCAL PACE5000 Bedienungsanleitung Seite 12

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
3.
+ve supply port.
4.
-ve supply port.
5.
Maximum Working Pressure (MWP).
6.
Output port.
7.
Vent port.
8.
Reference port.
9.
PACE5000
10. Control module.
11. PACE6000
12. Blanking plates.
1.1 Pressure Adaptors
Figure B1 shows the available range of PACE pressure adaptors. Refer to Table 1 and the Data
Sheet for more information.
Adaptor Part Number
IO-SNUBBER-1
IO-DIFFUSER-1
IO-ADAPT-1/4NPT
IO-ADAPT-1/8NPT
IO-ADAPT-7/16UNF
IO-ADAPT-AN4
IO-ADAPT-AN6
IO-ADAPT-BARB
IO-ADAPT-G1/4
IO-ADAPT-G1/8
1.2 Pressure Connection
WARNING
thread to ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
Tapered threads not allowed.
The PACE has parallel thread pressure connectors. Use only the connector type specified in
Table 2.
Table 2: PACE Pressure Connector Thread Specification
PACE Connector
Supply +, Supply -, Output, Vent,
Reference
Refer to Figure B2 for connection to the PACE pressure connectors and the key below:
1.
PACE pressure connector.
2.
Bonded seal.
Copyright 2008 Baker Hughes Company.
4 | PACE5000/6000 Safety Instructions–English
Table 1: Pressure Adaptor Specification
Specification
Restrictor/Snubber
Diffuser
ISO 228 G1/8 Male to 1/4 NPT Female.
ISO 228 G1/8 Male to 1/8 NPT Female.
ISO 228 G1/8 Male to 7/16-20 UNF Female.
ISO 228 G1/8 Male to AN4 37° Male.
ISO 228 G1/8 Male to AN6 37° Male.
ISO 228 G1/8 Male to 1/4 Hose.
ISO 228 G1/8 Male to ISO 228 G1/4 Female.
ISO 228 G1/8 Male to ISO 228 G1/8 Female.
Parallel threads must be used. Female thread type is parallel
ISO228/1 G1/8 Parallel Threads (DIN ISO228/1, JIS B0202)
Thread Specification