3.
ISO228/1 G1/8 pressure connector.
4.
NPT thread pressure connector.
5.
Pressure adaptor, see "Pressure Adaptors" on page 4.
For pressures less than 100 bar (1450 psi), see alternative sealing method in Figure B3 and the
key below:
1.
PACE pressure connector.
2.
Bonded seal.
3.
ISO228/1 G1/8 pressure connector or adaptor. For adaptors, see "Pressure Adaptors" on
page 4.
1.3 Pressure Supply
Refer to Figure B4 for connection of the PACE pressure supply and the key below:
1.
Connector
2.
Bonded seal
Note: For NPT connections use a suitable thread adaptor. Refer to "Pressure Adaptors" on
page 4 for details.
The pressure supply must be clean, dry, nitrogen or air and at the correct pressure (refer to
the supply equipment below).
Ensure the user systems can be isolated and vented.
Connect pressure and vacuum supplies to the SUPPLY + and SUPPLY – connection ports.
Connect the Unit Under Test (UUT) to the required output connection port.
2. Installation
The instrument requires a positive pressure supply, instruments operating in an absolute range
or negative pressure range require a vacuum supply. A vacuum supply should be used for a fast
response for instruments operating near atmospheric pressure. For dual channel operation two
independent pressure and vacuum supplies can be used.
Note: When using two pressure modules (Refer to K0443, Section 2.6, System), make sure that:
The module with the highest pressure rating is fitted to the right hand side Module 1 position
as viewed from the rear of the product, refer to (Refer to Figure A1)
If two modules have the same pressure rating, make sure that the module with the higher
serial number is fitted to the right hand side Module 1 position as viewed from the rear of the
product.
Note: All pneumatic connections must comply with the Pressure Equipment Directive (PED) or
other regional pressure standards.
Note: When connecting the output ports of two pressure modules together make sure both are
either:
below 70 bar (1015 psi)
OR
between 100 to 210 bar (1450 to 3045 psi).
Copyright 2008 Baker Hughes Company.
English–PACE5000/6000 Safety Instructions | 5