ASEKO CLF Probe
The ASEKO CLF probe is intended for measurement of free chlorine content in
water. The unique structure of the membrane probe with the built-in electronics
makes it possible to use the free chlorine probe even under conditions where
other probes cannot be used. The probe is designed to be screwed into a fl ow
well being part of ASIN Aqua automatons controlling pool water quality.
It is delivered consisting of:
• Probe
• Membrane module
• Electrolyte bottle
8
Commissioning
1. Unscrew the membrane module.
2. Fill the module with electrolyte up to a mark (see "Electrolyte Filling").
3. Tap the module to remove possible air bubbles.
4. Easy screw the electrode into the module while keeping the module with
a hole up and holding its lower part (membrane) with a fi nger (the fi nger
must be clean, free of grease) so as not to break or deform it.
5. An excess electrolyte will slowly fl ow o through a miniature hole in a part
where the module is threaded.
6. Before inserting the probe into the automaton fl ow well, shake the probe
similarly as shaking o a clinical thermometer (always with the connector
up).