screwing

Probe Maintenance

In case of outdoor pools being put out of service through the winter season,
it is necessary to remove the probe from the well and unscrew it from the
membrane module, rinse all with clean water (be aware that water must not
enter the connector) and store in a dry place at room temperature. While
the probe is in operation, its calibration can fail or the probe starts showing
error data. In this case it is necessary to replace electrolyte, the membrane
module or the entire probe.
Electrolyte Replacement
Proceed similarly as in the case of commissioning: Unscrew the
membrane module. Clean the probe tip and body using a paper and do
not touch it any more. Rinse the membrane module with distilled water.
While commissioning, proceed from point 2. Unless the situation is
remedied, the membrane is probably damaged and it is necessary to order
a new membrane module, refurbish or replace the probe with a new one.