Hold the meter at a right angle and to the centre of the fly trap when taking a
measurement, see figure 1.
TABLE 2
LED Number
1
2
LED Colour
RED
Lamp(s)
INADEQUATE
Condition
LAMPS SHOULD BE
REPLACED
Note – Slight movements away from this
angle will give incorrect readings
The 10-point coloured LED bar indicator
scale is used to determine the condition of
the UVA output levels of a given trap. Table
2 below explains how the scale should be
interpreted.
3
4
5
6
7
YELLOW
MARGINAL OUTPUT
GOOD
APPROACHING
SATISFACTORY
RENEWAL
OPERATING THE METER
1. Using Table 1 establish the measuring distance point from the fly trap.
2. Press either button labelled 'Amb' (Ambient) or 'Trap' to turn the meter on – The
LED's will illuminate in a sweeping motion form left to right and back again,
twice. After the sweeping the red LED number 1 will blink on and off repeatedly.
3. To measure the ambient UV light level you must first switch OFF the fly trap.
Point the meter in the general direction of the fly trap at the distance as specified
and press the 'Amb' button. The green LED number 10 will now blink on and
off repeatedly. This indicates that the meter has stored the ambient UV level
of the room.
4. Switch ON the fly trap.
5. Point the meter directly at the UV lamps in the fly trap at the correct distance
8
9
10
and press the 'Trap' button once, See figure 1.
6. The meter will now have measured the UV light levels emitting from the trap
GREEN
and will have subtracted the ambient UV level in the room. The result will then
be displayed on one of the 10 LED bar indicators by illuminating a single LED
continuously. Note – the red LED number 1 will also blink on and off repeatedly
to indicate the meter has performed its calculation correctly.