Herunterladen Inhalt Inhalt Diese Seite drucken

Specifications; Using The Thermometer - Eks 8009 Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 9
GB
Caution! Read the instructions before use!

Specifications

1. Temperature measurement range: 34-42.2°C (93.2~108°F)
2. Operating temperature range: 10~40°C (50~104°F)
3. Storage temperature range: -20~50°C (-4~122°F), RH≤85%
Transportation temperature should be less than 70°C, RH≤95%
4. Compliance with ASTM E1965-98, EN12470-5:2003 Clinical thermometers -
Part 5:Performance of infra-red ear thermometers (with maximum device),
IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safety) standards.
5. This mode converts the ear temperature to display its "oral equivalent. " (according to the
result of the clinical evaluation)
6. °C / °F switch function
7. Battery: one lithium battery CR2032
8. Auto power shut-off
9. Accuracy:+/-0.2°C (0.4°F), 35.5~42°C (95.9~107.6°F); +/-0.3°C (0.5°F) for other range.
2

Using the thermometer

1. Gently squeeze the sides of the probe cap and twist to
remove. Do not use force in removing the cap.
2. Always use a new and undamaged probe cover.
Attach the probe cover to the connection ring. Fig.1
Important: Place the smooth side (not the adhesive
• Place the center of the probe cover over the end of the probe.
• Insert and push the connection ring until the probe cover
clicks into place. The "click" means that the probe cover has
been connected firmly. Fig. 2
Warning: Proper installation of the probe cover ensures accurate
measurements. If the probe cover has not been connected firmly,
the probe cover icon
please check the setting of the probe cover again.
Warning: Keep the probe covers and connection ring away from
children.
3. Make sure the ear passage is clean.
4. Press the ON/MEM button. The thermometer is ready
for use after the ear icon stops flashing and two beeps sound.
5. Gently pull the ear back and up to straighten the ear passage. Fig. 3
To obtain the most accurate reading snugly position the
probe in the ear so that the probe point is in line with the
membrane of the eardrum. Fig. 4
6. With the index finger press the SCAN button until you hear a
beep. Fig. 5
7. The device will shut off automatically if not used for more than 1
minute. This saves battery life.
3
Fever alarm:
If the thermometer detects a body temperature of more than
37.5°C when measuring by the ear or forehead, there will be a
long beep sound followed by three short beep sounds which
warn of potential fever.
4
Memory - MEM
If the temperature reading is within the normal range of 34 °C to
42.2 °C (93.2 °F to 108.0 °F) when the thermometer is turned off,
the last measurement is saved in the memory.
To recall the previous temperature press the ON/MEN button
once again.
will flash on the display. In this case

Fig.1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis