Slings

Care Instructions

Check for wear and tear
Check the sling for wear and tear. It is important
to check the fabric, the edging, the seams and
the lifting straps.
Damaged lifting equipment must not be used.
Cleaning
Human Care slings are intended for individual
use. To avoid infections or transmission of bac-
teria, it is important to clean the slings properly.
All slings should be machine washed and dried
as soon as possible. Slings should always be
cleaned after use in a wet situation.
Always refer to the label of the sling before
washing.

Periodic inspection

A periodic inspection must be carried out at
least every 6 months. See Human Care's peri-
odic inspection log which can be found on our
website at www.humancaregroup.com
14
www.humancaregroup.com

Product lifetime

The product´s expected lifetime 3 years or
150-180 times washing, if the product is used
as intended and maintained according to the
manufacturer´s instructions, depending on the
intensity of use and maximum load applied
during use. If the product label is no longer
legible, the product should be discarded.
Warranty and support
If you need information or support, please
contact your Human Care office,
www.humancaregroup.com, or distributor.
Recycling
Once the product has reached the end of its life,
it should be disposed according to your coun-
try's rules, especially for the separate collection
of electrical and electronic products.
Correct disposal helps prevent negative
consequences for the environment and human
health.