Herunterladen Inhalt Inhalt Diese Seite drucken

Venen Engel 6 PREMIUM Bedienungsanleitung Seite 42

Vorschau ausblenden Andere Handbücher für 6 PREMIUM:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in
such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4-11
Power frequency (50Hz/60Hz)
magnetic field IEC 61000-4-8
Note: Ur is the AC mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
±8 kV contact
±15 kV air
±2 kV for power supply lines
± 1 kV line(s) to line(s)
<5% UT (>95% dip
in UT) for 0.5 cycle 40% UT (60% dip
in UT) for 5 cycles 70%UT(30%
dip in UT) for 25 cycles <5% UT
(>95% dip in UT) for 5 sec
l0V/m
Compliance level
±8 kV
Floors should be wood, concrete or ceramic tile. lf
contact ±15 kV air
floor are covered with synthetic materi- al, the relative
humidity should be at least 30%.
±2kV for power supply
Mains power quality should be that of a typical
lines
commercial or hospital environment.
±1 kV
Mains power quality should be that of a typical
commercial or hospital environment.
differential mode
<5% UT (>95%
Mains power quality should be that of a typical
commercial
dip in UT) for
or hospital environment. lf the user of the DEVICE
0.5 cycle 40% UT (60%
requires continued operation during power mains
dip in UT) for 5 cycles
interruptions,
70% UT (30% dip in
it is recommended that the DEVICE be powered
UT) for25
from an uninterruptible power supply or a battery.
cycles <5% UT
(>950/odip in UT) for 5
sec
l0V/m
Power frequency magnetic field should be at levels
characteristic of a typical location in a typical
commercial or hospital environment.
Electromagnetic environment
- guidance
42

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für Venen Engel 6 PREMIUM

Diese Anleitung auch für:

4

Inhaltsverzeichnis